Table 2.

Recommendations for design and implementation of PBRTQC laboratory protocols

Subject Recommendation
Management of PBRTQC settings When PBRTQC is used, the appropriateness of the settings has to be guarded in a proactive manner to ensure and maintain an acceptable false alarm rate
PBRTQC alarm actions PBRTQC alarm actions to be considered include:
- Stop result reporting from the affected instrument
- Move samples to an alternative instrument
- Determine whether an analytical error occurred and, if so, determine the size of the error
- Initiate troubleshooting protocol
PBRTQC alarm workup PBRTQC alarm workup steps to be considered are:
- Analyze internal quality control
- Reanalyze recently measured samples on alternative analyzer
- Run external quality control
- Review recently obtained results from affected assay
Recovery from PBRTQC alarm Reset/restart PBRTQC calculations after PBRTQC workup completion

Adapted from the article of Badrick et al. (Clin Chem. 2019;65:962-971) [25] with original copyright holder’s permission.

Abbreviation: PBRTQC, patient-based real-time quality control.

Korean J Clin Lab Sci 2024;56:105-14 https://doi.org/10.15324/kjcls.2024.56.2.105